Turbidity meter, Handheld, Infrared Light, 0-1100 NTU

QT20008
Turbidity meter, Handheld, Infrared Light, 0-1100 NTU


Product number     QT20008
OEM reference     20008

  • Auto ranging 0 - 1100 NTU
  • Instrument senses turbidity level of sample and automatically adjusts to the appropriate measurement
  • Rugged portable carrying case
  • Ensures durability and convenience no matter where your sample location happens to be
  • Simple calibration procedures
  • Calibration initiated with the push of a button ensures accurate readings
  • Waterproof housing allows sample measuring and cleaning in any wet environment
  • Completely self-contained
  • Contains everything you need, including battery pack, manual, and calibration standards.
  • Long battery life. Over 5000 Tests on a single set of 4 - AAA alkaline batteries
  • ProCal calibration standards. Quick and easy calibration across all ranges without the need to mix Formazin


The continuous monitoring NTU system shall include the online model with power supply, display, user interface and sensor as one single SKU. It will not be necessary to purchase a separate module for display or power supply. The turbidimeter shall meet all requirements specified by the USEPA Method 180.1 (White
Light Model), ISO 7027 (Infrared Model) and Standard Methods 2130B. The turbidimeter shall be Modbus compatible and have a 4-20 mA output. The turbidimeter shall have the same optical design as the M100+ laboratory turbidimeter.

The online turbidimeter shall have an ultrasonic auto-clean system. Accuracy shall be 2% of reading or plus or minus 0.02, whichever is greater, from 0-40 NTU, and 5% of reading or plus or minus 0.02, whichever is greater, from 40-1000 NTU. Resolution will be 0.001 NTU (user selectable).

The sensor shall utilize a rotational flow through assemby. The sensor shall utilize a rotational flow through assembly with a 30ml measuring cuvette. The specially designed flow head bubble rejection system will reduce the need for a bubble trap and provide a faster response time. The sample stream will remain under pressure, ensuring it will not be exposed to open atmosphere, which can cause precipitants in some circumstances. Precipitation of oxidized sample constituents can result in false NTU readings. The sensor shall allow grab samples to be measured without the need to stop sample flow to the flow head.

Calibration and standardization will be accomplished using small volumes (30ml) of reusable primary standards. Primary Standards shall be reusable for multiple online turbidimeters and also with the M100+ laboratory turbidimeter. Calibration procedures can be completed without disrupting the sample flow. Details of calibration (date and time of last successful calibration) will be stored on the analyzer. An integral part of the turbidimeter system is the inclusion of a laboratory instrument specifically designed to use the same calibration standards. The required supporting laboratory instrument must be an HF scientific M100+ touchscreen model with USB data logging and ability to store the date/time of the last calibration.

The process turbidimeter and the laboratory turbidimeter lamp source and detector shall not come in contact with the sample, eliminating false low readings. The turbidimeters shall use menu driven software. The online turbidimeter enclosure shall be ABS. The online turbidimeter will have data logging features, will store the latest date/time of calibration and be user selectable for ranges of 0 – 10 NTU, 0 – 100 NTU (factory default) or 0 – 1000 NTU. The online turbidimeter shall be HF scientific MTOL+ Online Turbidimeter.
  • Conformance ISO7027 Method (IR)
  • Range Auto ranging from 0 - 1100 NTU
  • Resolution 0.01 NTU < 100 NTU
  • 0.1 NTU < 100 - 999.9 NTU
  • 1 NTU < 1000 - 1100 NTU
  • Accuracy ± 2% of reading or ± 0.01 NTU (0-500 NTU)
  • ± 3% of reading (500-1100 NTU)
  • Response time 14 seconds
  • Power supply 4 - AAA alkaline batteries (over 5000 tests)
  • Display 4 digit (7 segment) LCD
  • Principle of operation Nephelometric
  • Certification CE, NEMA 4X. Designed to meet IP67
  • Light source Infrared LED (860nm)
  • Sample cells 10 ml