Dissolved oxygen electrode, 316SS, Ø25x320mm
QD2532X
Dissolved oxygen electrode, 316SS, Ø25x320mm
Dissolved oxygen electrode, 316SS, Ø25x320mm
Product number | QD2532X |
Application type |
|
- Polarisation voltage approx 680mV
- Output 25-50nA
The steam sterilizable family of dissolved oxygen electrodes are specifically designed for fermentation where it is necessary to subject electrodes to repeated steam sterilization. Dissolved oxygen electrodes are available in both polarographic and galvanic styles to fulfill all requirements. Van London Co. O2 electrodes are compatible with most common O2 amplifiers found in fermentation.
Polarographic electrodes are designed to be used with Van London Co., Ingold IL, Cole-Parmer and Applikon O2 amplifiers, or with other amplifiers that are designed to be used with Ingold dissolved oxygen electrodes.
Dissolved Oxygen probes are available for laboratory fermentation where sterilization takes place in a separate autoclave or industrial applications where sterilization takes place in-situ. The easily replaced electrode membrane cartridge is a unique feature of the polarographic and NBS galvanic dissolved oxygen electrodes. The membrane is of a special construction - a silicone and teflon composite, reinforced with stainless steel mesh. This unique membrane is very rugged and allows for repeated sterilization without damage, allowing it to withstand severe conditions. The electrode operates without maintenance for 5 to 10 sterilizations, fermentation cycles or, in terms of time, for 6 to 8 weeks.
Polarographic electrodes are designed to be used with Van London Co., Ingold IL, Cole-Parmer and Applikon O2 amplifiers, or with other amplifiers that are designed to be used with Ingold dissolved oxygen electrodes.
Dissolved Oxygen probes are available for laboratory fermentation where sterilization takes place in a separate autoclave or industrial applications where sterilization takes place in-situ. The easily replaced electrode membrane cartridge is a unique feature of the polarographic and NBS galvanic dissolved oxygen electrodes. The membrane is of a special construction - a silicone and teflon composite, reinforced with stainless steel mesh. This unique membrane is very rugged and allows for repeated sterilization without damage, allowing it to withstand severe conditions. The electrode operates without maintenance for 5 to 10 sterilizations, fermentation cycles or, in terms of time, for 6 to 8 weeks.
Dissolved Oxygen electrode | QD2532X |
Range | 0..200% - 0..20mg/l |
Temperature range (°C) | 0..80 |
Shaft material | 316 SS |
Shaft diameter (mm) | 25 |
Shaft length (mm) | 320 |
Output | 50..60 nA at air (25 °C, 1013 mbar) |
Stability | Better then 2% |
Residual signal | < 1 nA |
Sterilisation Temperature (°C) | 130 |
Temperature compensation | yes |
Connector | 4-pin |
ATC | NTC 22kOhm |